High
Vacuum Pumping System VPC-250F
|
|
VPC series / Oil diffusion pump (air cooling) + Oil rotary vacuum pump
|
 |
|
| Outline |
It is a small type high vacuum pumping system that necessary parts are compactly placed in a pipe-framed stand. It is easy to handle, maintain, and suitable for experiment. |
| |
| Feature |
| 1. |
It is easily moving vacuum pumping system with casters. |
| 2. |
The vacuum chamber can be opened without stopping the main pump. |
| 3. |
Maintenance is easy, because it is made of few parts included a three way valve. |
| 4. |
Cooling water of an oil diffusion pump is unnecessary because of an air cooling type. |
|
|
| Model |
VPC-250F |
| Ultimate
pressure |
10-4Pa
10-5Pa (with LN cold trap) |
| Main pump |
Oil diffusion pump (air cooling) 200 L/sec |
| Backing pump |
Oil rotary vacuum pump 100 L/min |
| Main valve |
65A Clapper valve |
| Sub valve |
Three way valve |
| Inlet flange |
VG-65 |
| Weight |
55.7kg |
| Overall dimensions |
530mm(W)×537mm(D)×735mm(H) |
|
* Ionization vacuum gauge · Pirani vacuum gauge · LN cold trap · Conversion flange · Oil mist trap are options. |
top of page
Model |
VPC-250F |
| Power required |
Single phase, 50/60Hz, 100V, 1.35kVA |
| Ground terminal |
D grade (ground resistance/100MΩ or less) |
| Connection |
Vinyl cabtyre cable (with a plug) 2m |
|
top of page
 |
Model |
DPF-200 |
| Ultimate
pressure |
10-5Pa |
| Maximum evacuation capacity |
200L/sec |
| Critical back-pressure |
20Pa |
| Required power |
Single phase, 50/60Hz, 100V, 0.45kW |
| Recommended oil |
SX 70cc |
| Weight |
5kg |
| Inlet flange |
VG-65 |
| Outlet flange |
dia. 27mm |
| Cooling method |
air cooling |
|
* You can see each option, if you click an option. |
|

top of page
ULVAC KIKO, Inc. |
Overseas Division |
TEL (81) 45-533-0206 |
FAX (81) 45-533-0204 |
Please contact us by E-mail → here |
|
|